Transmitter SITRANS F C MASSFLO MASS 6000 Compact IP67/NEMA 4X | Measurement of | Mass flow [kg/s (lbs/min)], volume flow [l/s (gpm)], fraction [%], °Brix, density [kg/m3, (lbs/ft3)], temperature [°C (°F)] | Current output | | | 0 ... 20 mA or 4 ... 20 mA | | < 800 Ω | | 0 ... 99.9 s adjustable | Digital output | | | 0 ... 10 kHz, 50% duty cycle | | 0 ... 99.9 s adjustable | | 24 V DC, 30 mA, 1 KΩ ≤ Rload ≤ 10 KΩ, short-circuit-protected | | 3 ... 30 V DC, max. 110 mA, 1 KΩ ≤ Rload ≤ 10 KΩ | Relay | | | Change-over relay | | 42 V/2 A peak | | Error level, error number, limit, flow direction | Digital input | 11 ... 30 V DC (Ri = 13.6 KΩ) | | Start/hold/continue batch, zero point adjust, reset totalizer 1/2, force output, freeze output | Galvanic isolation | All inputs and outputs are galvanically isolated, isolation voltage 500 V | Cut-off | | | 0 ... 9.9 % of maximum flow | Limit function | Mass flow, volume flow, fraction, density, sensor temperature | Totalizer | Two eight-digit counters for forward, net or reverse flow | Display | - Background illumination with alphanumerical text, 3 × 20 characters to indicate flow rate, totalized values, settings and faults. Time constant as current output 1
- Reverse flow indicated by negative sign
| Zero point adjustment | Manual via keypad or remote via digital input | Ambient temperature | | | -20 ... +50 °C (-4 ... +122 °F), max. rel. humidity 80% to 31 °C (87.8 °F) decreasing to 50% at 40 °C (104 °F) according to IEC/EN/UL 61010-1 | | -40 ... +70 °C (-4 ... +158 °F) (Humidity max. 95 %) | Communication | Add-on modules: HART, PROFIBUS PA & DP, MODBUS RTU RS 485, DeviceNet, FOUNDATION Fieldbus H1 | Enclosure | | | Fibre glass-reinforced polyamide | | IP67/NEMA 4X to IEC 529 and DIN 40050 (1 m w.g. for 30 min.) | | 18 ... 1000 Hz random, 3.17G rms, in all directions, to IEC 68-2-36 | Supply voltage | | 24 V version | | | 24 V DC/AC, 50 ... 60 Hz | | 18 ... 30 V DC 20 ... 30 V AC | | 10 W | 230 V version | | | 87 ... 253 V AC, 50 ... 60 Hz | | 26 VA | Fuse | | | T400 mA, T 250 V (IEC 127) - Not to be changed by user | | T1 A, T 250 V (IEC 127) - Not to be changed by user | EMC performance | | | EN/IEC 61000-6-4 (Industry) | | EN/IEC 61000-6-2 (Industry) | NAMUR | Within the value limits according to "Allgemeine Anforderung" with error criteria A in accordance with NE 21 | Environment | | - Environmental conditions acc. to IEC/EN/UL 61010-1:
| - Altitude up to 2000 m
- POLLUTION DEGREE 2
| Maintenance | The flowmeter has a built-in error log/pending menu which should be inspected on a regular basis | Cable glands | Two types of cable gland are available in polyamide in the following dimensions: M20 or ½” NPT |
|
|
|
|
|
MASS 6000 for 19" insert/19" wall mounting
|
|
Mã sản phẩm:
7ME4110
Tên sản phẩm:
MASS 6000 for 19" insert/19" wall mounting
SITRANS F C MASSFLO MASS 6000 19” IP20/NEMA 1 or IP66/NEMA 4 version | Measurement of | Mass flow [kg/s (lbs/min)], volume flow [l/s (gpm)], fraction [%], °Brix, density [kg/m33)], temperature [°C (°F)] (lbs/ft | Current output | | | 0 ... 20 mA or 4 ... 20 mA | | < 800 Ω | | 0 ... 99.9 s adjustable | Digital output | | | 0 ... 10 kHz, 50 % duty cycle | | 0 ... 30 s adjustable | | 24 V DC, 30 mA, 1 KΩ ≤ Rload ≤ 10 KΩ, short-circuit-protected | | 3 ... 30 V DC, max. 110 mA, 1 KΩ ≤ Rload ≤ 10 KΩ | Relay | | | Change-over relay | | 42 V / 2 A peak | | Error level, error number, limit, direction | Digital input | 11 ... 30 V DC | | Start/hold/continue batch, zero point adjust, reset totalizer 1/2, force output, freeze output | Galvanic isolation | All inputs and outputs are galvanically isolated, isolation voltage 500 V | Cut-off | | | 0 ... 9.9 % of maximum flow | Limit function | Mass flow, volume flow, fraction, density, sensor temperature | Totalizer | Two eight-digit counters for forward, net or reverse flow | Display | - Background illumination with alphanumerical text, 3 × 20 characters to indicate flow rate, totalized values, settings and faults
- Reverse flow indicated by negative sign
| Zero point adjustment | Manual via keypad or remote via digital input | Ambient temperature | | | -20 ... +50°C (-4 ... +122 °F) | | -40 ... +70°C (-40 ... +158 °F) (Humidity max. 95 %) | Communication | Add-on modules: HART, PROFIBUS PA and DP, MODBUS RTU RS 485, DeviceNet, FOUNDATION Fieldbus H1 | Enclosure 19” | | | Aluminium/steel (DIN 41494) | | IP20/NEMA 1 to IEC 529 and DIN 40050 (1 m w.g. for 30 min.) | | 18 ... 1000 Hz random, 3.17G rms, in all directions, to IEC 68-2-36 | Supply voltage | - 87 ... 253 V AC +10% ... -10%, 50 ... 60 Hz
- 18 ... 30 V DC or 20 ... 30 V AC
| Power consumption | | | 9 VA max. | | 6 W IN = 250 mA, IST = 2 A (30 ms) | EMC performance | | | EN/IEC 61000-6-4 (Industy) | | EN/IEC 61000-6-2 (Industy) | Ex approval | [EEx ia] IIC, DEMKO 03 ATEX 135251X | Maintenance | The flowmeter has a built-in error log/pending menu which should be inspected on a regular basis | Fuse | T 400 mA, T 250 V (IEC 127), not replaceable by operator | Cable | - Max. 300 m
- C: max. 300 [pF/m]; LC/RC: max. 100 [μH/Ω]
- The total cable capacity must be max. 200 nF
| Cable glands | The cable gland is available in polyamide, in dimension: PG 13.5 |
|
|